1、DENSITOMETERCALIBRATIONPROCEDURENDE07-06Revision0Date30/04/2017Page6of6DENSITOMETERCALIBRATIONPROCEDURE3210REVISIONPREPAREDBYREVIEWEDBYAPPROVEDBYAIDENSITOMETERCALIBRATIONPROCEDURENDE07-06Revision0Date30/04/2017Page6of61.ScopeThisprocedurecoversthecalibrat
2、ionoftransmissiondensitometersusedtoperformradiographicfilmdensitymeasurementsandprovidesameansofevaluatingthereliabilityoftransmissiondensitometersusedforthemeasurementofradiographicfilmdensity.ThisprocedureispreparedreferencetoT262"DENSITOMETERANDWEDGEC
3、OMPARISONFILM”ofASMESect.Ⅴ.2.Apparatus2.1Apparatusshouldconsistofthefollowing:2.1.1TheX-rayfilmsteptabletstandardoftheNationalCenterofMeasuringhavingatleast5stepswithneutraldensitiesfromatleast1.0through4.0.Thecalibrationofthestepwedgefilmshallhavebeenver
4、ifiedwithinthelast12monthbycomparisonwithaNationalCenterofMeasuringserializedsteptablet.2.1.2TransmissionDensitometerswithdigitalreadoutdisplaysspecificallymanufacturedforthepurposeofmeasuringtherangeoffilmdensities,describedin2.1.1shallbeused.2.1.3Manufa
5、cturer'sOperatingInstructionforAppropriateTransmissionDensitometer.DENSITOMETERCALIBRATIONPROCEDURENDE07-06Revision0Date30/04/2017Page6of63.Calibration3.1Full-scalelinearitycalibrationofdensitometershallbeperformedatleastevery90daysduringuse.3.1.1Allowami
6、nimumof30min"warm-up"time.Afterthiswarm-upperiod,adjustthe"0"readingofthedensitometer.3.1.2Selectandpositionforreadingtheneutraldensityclosestto1.0,2.0,3.0,4.0onthecalibratedsteptablet.Readandrecordthedensityforeachstep.3.1.3Comparethedensitiesrecordedwit
7、htheactualdensityvalueslistedonthecalibratedorthedensityvaluelistedonthecalibrationcertificate.Ifthedensitometerhasbeencalibratedproperly,thedensitiesat1.0,2.0,3.0,4.0shouldnotvarymorethan±0.05densityunits.Ifanyoftherecordeddensityvarymorethan±0.05density
8、unitsfromthedensityvaluesonthecalibratedsteptablet,thelinearityofthedensitometerisoutoftoleranceandshallbecorrected.4.RecordsandAssociatedDocumentation4.1Thedensitometerrequiredby3.1shouldbenotedandrecordedthecalibr